ST140 has been designed to offer you pH, conductivity, total dissolved solids and temperature measurements all in a slim, lightweight, portable unit. From purified to brine water measurement, you can select the meter which will work best with your range of conductivity for greater precision.
There are only 2 buttons, yet you can select from a range of calibration buffers and even the temperature scale (°C or °F) most familiar to you. The housing is waterproof and rated for IP 67 conditions.
The ST1288 (Electrode) pre-amplified multiparameter probe features an easy to clean sensor and a cloth junction that can be pulled longer to increase the life of the probe. To ensure against interference from transient electrical noise, a solid-state amplifier is integrated into the ST1288(Electrode).
User selectable features include different TDS factors from 0.45 to 1.00 and a range of temperature coefficients (β) from 0.0 to 2.4% for greater consistency and reproducibility. Standardized buffer recognition values are also selectable.
This instrument easily fits in the palm of your hand and the bottom probe connection ensures the electrode cable doesn’t get in your way. The large, multi-level LCD displays the primary reading, temperature and calibration guides simultaneously.
Range |
pH |
0.00 to 14.00 pH |
|
EC |
0 to 3999 µS/cm |
|
TDS |
0 to 2000 ppm (mg/L) |
|
Temperature |
0.0 to 60.0°C (32.0 to 140.0°F) |
Resolution |
pH |
0.01 pH |
|
EC |
1 µS/cm |
|
TDS |
1 ppm |
|
Temperature |
0.1°C (0.1°F) |
Accuracy |
pH |
±0.01 pH |
|
EC |
±2% F.S. |
|
TDS |
±2% F.S. |
|
Temperature |
±0.5°C (±1.0°F) |
TDS Conversion Factor |
|
selectable 0.45 to 1.00 with 0.01 increments (default 0.50) |
pH Calibration |
|
automatic, one or two point calibration with two sets of memorized buffers (Standard 4.01, 7.01, 10.01 or NIST 4.01, 6.86, 9.18) |
EC/TDS Calibration |
|
automatic, 1 point at 1382 ppm (CONV 0.5) |
Temperature Compensation |
pH |
Automatic |
|
EC/TDS |
Automatic with selectable ß from 0.0 to 2.4%/°C with 0.1 increments |
Probe |
|
ST1288 PVC body, pre-amplified multiparameter probe with internal temperature sensor, DIN connector and 1.2 m (3.93’) cable (included) |
Battery Type / Life |
|
1.5V AAA (3) /approximately 500 hours of continuous use. auto-off after 8 minutes of inactivity |
Environment |
|
0 to 50°C (32 to 122°F); RH max 100% |
Dimensions |
|
152 x 58 x 30 mm (6.0 x 2.3 x 1.2”) |
Weight |
|
205 g (7.2 oz.) |
Comment :