Wide wavelength range, satisfying requirements of various fields
Five options for spectral bandwidth selection, 5nm, 4nm, 2nm, 1nm and 0.5nm, made according to customer's need and satisfying the requirements of pharmacopoeia
Fully automated design, realizing the simplest measurement
Optimized optics and large scale integrated circuits design, light source and receiver from world famous manufacturer all add up to high performance and reliability
Rich measurement methods, wavelength scan, time scan, multi-wavelength determination, multi-order derivative determination, double-wavelength method and triple-wavelength method etc., meet different measurement requirements
Automatic 10mm 8-cell holder, changeable to automatic 5mm-50mm 4-position cell holder for more choices
Data output can be obtained via a printer port and a RS-232 interface (RS-485 and USB port optional)Parameters and data can be saved for user's convenience
PC controlled measurement can be achieved for more accurate and flexible requirement
Wavelength Range: 190-1100nm
Spectral Bandwidth: 2nm (5nm, 4nm, 1nm, 0.5nm optional)
Wavelength Accuracy: ±0.3nm
Wavelength Reproducibility: ≤0.15nm
Optical System: Double beam
Photometric Accuracy: ±0.3%T(0-100%T), ±0.002A(0-1A)
Photometric Reproducibility: ≤0.15%T
Working Mode: T, A, C, E
Photometric Reproducibility: -0.3-3.5A
Stray Light: ≤0.05%T(NaI, 220nm, NaNO2 340nm)
Baseline Flatness: ±0.002A
Stability: ≤0.001A/h ( at 500nm, after warming up)
Noise: ±0.001A (at 500nm, after warming up)
Display: 6 inches high light blue LCD
Detector: silicon photo-diode
Power: AC 220V/50Hz, 110V/60Hz, 180W
Dimensions: 630×470× 210 mm
Weight: 26kg