●Occupy less place, Simple Operation
●Easy-to-read LCD Display with high luminant backlight
●Unit display: mg/L or %
●One key to check through all the settings, including: Zero offset, slope and all the settings
●Clark cell mode, long lifespan
●256 data sets memory stores and recalls
●Auto Power off saves battery after 10-mins non-use
●Detachable Electrode Stand organizes multiple electrodes neatly, easy installing at either left or right side and holds them firmly in place
Saturation Percentage |
Range |
0.0%~400.0% |
|
Resolution |
0.1% |
|
Accuracy |
0.2% F.S. |
Oxygen concentration |
Range |
0.00 mg/L~40.00 mg/L |
|
Resolution |
0.01 mg/L |
|
Accuracy |
0.2% F.S. |
Temperature |
Range |
when compensating:0.0 C~50.0 C |
|
Resolution |
0.1°C |
|
Accuracy |
0.5 C |
Atmospheric Pressure |
Range |
600 mbar~1400 mbar |
|
Resolution |
1 mbar |
|
Default |
1013 mbar |
Salinity |
Range |
0.0 g/L~40.0 g/L |
|
Resolution |
0.1 g/L |
|
Default |
0.0 g/L |
Environment |
Humidity |
-5°C~60°C, Relative humidity<90% |
Dimensions |
140*210*35mm (W*L*H) |
|
Weight |
630g |
Comment :