Equipment |
Meter with power cord, without sensor |
Ranges |
Cond: 0.000 ... 9.999 μS/cm; 00.00 ... 99.99 μS/cm; 000.0 ... 999.9 μS/cm; 0.000 ... 9.999 mS/cm; 00.00 ... 99.99 mS/cm; 000.0 ... 999.9 mS/cm; 0000 ... 2000 mS/cm; auto-ranging or manual preset °C: –50.0 ... +150.0 |
Display |
Alphanumer. 2 x 4-digit, 14-segment LED, character height 13 mm, measurement symbols:20 °C, 25 °C, μS/cm, mS/cm, %/K, °C, Time, 3 Sensoface icons inform on sensor and measuring equipment (GLP) |
Measuring cycle |
Approx. 1.5/sec |
Measuring frequencies |
Approx. 40 Hz to 2 kHz, automatic adjustment by conductance |
Resolution |
Up to 0.001 μS/cm |
Accuracy |
Cond: <0.5 % meas. value. ±2 counts °C: 0.3 K |
Reproducibility |
<0.1 % meas. val. |
Temperature compensation |
–50 ... +150 °C; Pt 1000/NTC 30 kOhm (autom. selection) or manual, Linear TC characteristic 0.00 ... +9.99 %/K, ref. temperature 20 °C/25 °C selectable |
Adm. cell constant |
0.001 ... 199.9 cm–, adjustable |
Sensor standardization |
Operating modes – Automatic by determining the cell constant with NaCl or KCl solution Calibration solutions: KCl 0.01 mol/l; 0.1 mol/l; 1 mol/l; NaCl 0.01 mol/l; 0.1 mol/l; saturated – Direct entry of cell constant |
Monitoring of sensor and equipment (GLP) |
Sensoface provides information: – for selection of 2-electrode sensors – on too great a difference between reference and measuring temperature – for handling of 4-electrode sensors – on clock memory loss – in case of irregular checking of measuring equipment – Optical display: good/average/poor |
Device self-test |
Test of measuring electronics including recorder output, segment and keypad test, RAM, EPROM, and EEPROM test during diagnostics, automatic short check at power-on |
GLP records (ISO 9000)
|
Parameter settings, calibration, device diagnostics |
Recorder output |
Galvanically isolated (isolation voltage: 40 V DC, 20 V AC) Cond: 1 mV/μS · cm–!; 1 mV/mS · cm–!; °C: 10 mV/°C user-defined for printer control |
Interface
|
RS 232 without control lines, galvanically isolated (isolation voltage: 40 V DC, 20 V AC), user-definable as printer or computer interface,Baud rate: 600 / 1200 / 2400 / 4800 / 9600 Data bits/parity: 7/Even, 7/Odd, 8/No Protocol: none, XON/XOFF Stop bits: 1 |
Software |
Control of the Model ST 700 Laboratory Conductivity Meter is integrated in the automation software for lab meters “labworldsoft” for display and control of device functions for Version 4.0 or higher |
Clock |
Real-time clock with date, self-contained |
Calibration data storage |
Automatic storage of cell constant and calibration procedure with time and date stamp, self-contained |
Data retention
|
Parameters, statistics, and factory settings: >10 years (EEPROM) Clock: reserve power >1 year (battery-backed) |
Protection against electrical shock |
Protective separation as defined in DIN 57100 / VDE 0100 Part 410 and DIN VDE 0106 Part 101, power supply against all other inputs and outputs, in accordance with the NAMUR recommendation “Extra-low voltage circuits with protective separation” |
EMC directive |
89/336/EEC |
Standards |
EN 61326 / VDE 0843 Part 20: 2002-3 |
Ambient temperature Storage and transport temp |
0 ... +45 ؛C –20 ... +70 ؛C |
Power supply |
230 V –15 % +10 %, 48 ... 62 Hz, <10 VA, Option 363: 115 V AC |
Sensor connection |
The meter allows connection of any 2-electrode sensors with banana plug.Special diode plug for 4-electrode sensors |
Enclosure |
Glass-reinforced polyamide 12, stainless steel cover, IP 54 protection, prepared for connecting ZU 6954 attachable stand |
Dimensions (W x H x D) |
244 x 95 x 255 mm |
Weight |
Approx. 2 kg |
Comment :